En el presente estudio se realizó una medición de gases arteriales y electrolitos en 50 caninos sanos en la sabana de Bogotá, particularmente en criaderos caninos de los municipios de Chía, Cajicá, Cota y Bogotá D.C., utilizando el equipo Epoc® y sus respectivos cartuchos BGE. Los caninos, todos con edades entre 1 año y 9 años, se dividieron en 5 grupos de acuerdo a su peso: 0 a 15 Kg, 16 a 25 Kg, 26 a 35Kg, 36 a 45 Kg y mayores de 45 Kg de manera que cada grupo consistiera de 10 sujetos. Las muestras de sangre provenientes de la arteria femoral fueron analizadas de acuerdo a las instrucciones del fabricante utilizando cartuchos BGE los cuales arrojan resultados de pH, presiones de oxígeno (pO2) y dióxido de carbono (pCO2), Sodio (Na+), Potasio (K+), Calcio ionizado (iCa++), hematocrito (Htc), CO2 total (tCO2), bicarbonato (HCO3) - , saturación de oxígeno (cSO2), hemoglobina (cHgb) y exceso de bases (BE). A los valores se les realizó un análisis estadístico de ANAVA al azar utilizando a los grupos de peso como la variable categórica, un nivel de significancia de P<0.05 y prueba de Tukey no planeada entre tratamientos significativos. También se realizaron comparaciones no estadísticas frente a literatura encontrada para contrastar valores de otros estudios frente a los del presente estudio. También se buscaron diferencias estadísticamente significativas entre los resultados de este estudio y uno realizado en una población similar pero mediante el equipo I-Stat® fabricado por la compañía Abbott Point of Care Inc. Según lo analizado no se encontraron diferencias significativas entre las categorías de grupos de peso para ninguna de las variables. Para electrolitos no se encontraron diferencias entre lo reportado en la literatura y esta investigación, sí se encontraron diferencias metodológicas para gases frente a valores previamente reportados y sí se encontraron diferencias significativas para la variable de PO2 entre esta investigación y una realizada en una población similar pero mediante el uso del equipo I-Stat®. Se concluye que los valores de gases sanguíneos no solamente son propios a la altura donde residen los caninos monitoreados, sino que también estos valores obtenidos, particularmente de PO2, pueden ser dependientes del equipo utilizado. Se recomienda obtener y utilizar rangos de referencia acordes a la altura donde se está ejerciendo el monitoreo y el equipo utilizado.

Resumen en lengua extranjera 1

In this study blood gases and electrolytes were measured in 50 healthy canines from kennels in the savanna of Bogotá, specifically from Chía, Cajicá, Cota and Bogotá D.C., by using an Epoc® device and its BGE cartridges. The dogs, with an age range between 1 and 9 years, were divided into 5 groups according to weight: 0 to 15 Kg, 16 to 25 Kg, 26 to 35Kg, 36 to 45 Kg and those heavier than 45 Kg, so that each group consisted of 10 dogs. The blood samples, which were taken from the femoral artery, were analyzed according to the manufacturer’s specifications with BGE cartridges which give values for pH, partial pressure of oxygen (pO2), partial pressure of carbon dioxide (pCO2), sodium (Na+ ), potassium K+ ), ionized calcium (Ca++), hematocrit (Hct), total carbon dioxide (cTCO2), bicarbonate (cHCO3 - ), base excess (BE), oxygen saturation (cSO2), and hemoglobin (cHgb). The results were analyzed with a random statistical ANAVA using the weight groups as a categorical variable with a significant level of P<0.05, and an unplanned Tukey test between significant treatments. Non-statistical comparisons were also done with other published values contrasting them to this study. Additionally this study was compared to a study with a similar population but which was performed with an I-Stat® device manufactured by Abbott Point of Care Inc. According to the analysis there were no statistically significant differences between the results of any of the age groups. In terms of electrolytes there were no differences between previously published values and those in this study. In terms of gases there were methodological differences between this study and other previous studies and there was a statistically significant difference between the pO2 from this study and that which was performed in a similar population but with an I-Stat device. It is concluded that not only are blood gas values dependent on the altitude where the monitored canines live, but also that these values may be device dependent; specially in terms of pO2. The recommendation is to obtain and use reference ranges which are proper to the altitude where they are being monitored and in accordance to the device used.

Palabras clave

Enfermedades en caninos, Enfermedades cardiacas en perros, Criaderos caninos

Tipo de documento

Trabajo de grado - Pregrado

Licencia Creative Commons

Creative Commons Attribution-Noncommercial-No Derivative Works 4.0 License
This work is licensed under a Creative Commons Attribution-Noncommercial-No Derivative Works 4.0 License.

Fecha de elaboración

2013

Programa académico

Medicina Veterinaria

Facultad

Facultad de Ciencias Agropecuarias

Publisher

Universidad de La Salle. Facultad de Ciencias Agropecuarias. Medicina Veterinaria

Compartir

COinS