Title

Performance of Random Dot stereoscopic test with Lea Symbols

Resumen

El examen de estereopsis informa sobre la calidad de la visión binocular del paciente y orienta sobre la existencia y severidad de supresiones binoculares. Objetivo: determinar la sensibilidad y especificidad de la prueba Random Dot con símbolos de Lea en niños de edad preescolar de la ciudad de Bucaramanga, entre agosto del 2009 y abril del 2010. Metodología: se aplicaron las pruebas a 349 niños. Se usó como referencia la prueba estereoscópica Randot® Preschool, confrontada con la nueva prueba Random Dot con símbolos de Lea. Dos observadores clínicos se encargaron de realizar cada uno una prueba diferente. Resultados: se observó una sensibilidad del 79,31 % y una especificidad del 80 % para la prueba Random Dot con símbolos de Lea. Valor Predictivo Positivo (VPP): 26,44 %; Valor Predictivo Negativo (VPN): 97,7 %. Conclusiones: la prueba Random Dot con símbolos de Lea no representa un valor confiable para usarla como única prueba de tamizaje visual en la detección de alteraciones de la visión binocular. La baja prevalencia (8,3 %) de la condición dentro de este estudio aumenta su VPN. Esto unido a su alta especificidad hace que la prueba sea muy eficiente cuando se trata de confirmar la probabilidad de no tener la alteración cuando otra prueba, previamente, la ha clasificado con resultados normales.

Palabras clave

sensibilidad, especificidad, prueba de estereoagudeza, valores predictivos

Abstract

The stereopsis test reports on the quality of the patient’s binocular vision and provides guidance on the existence and severity of binocular suppressions. Objective: To determine the sensitivity and specificity of the Random Dot test with Lea symbols in preschool children from the city of Bucaramanga, between August 2009 and April 2010. Methodology: The tests were applied to 349 children. The Randot® Preschool stereoscopic test was used as reference, confronted with the new Random Dot test with Lea symbols. Two clinical observers were responsible for performing a different test each. Results: A sensitivity of 79.31 % and a specificity of 80 % were observed for the Random Dot Test with Lea Symbols. Positive Predictive Value (PPV): 26.44 %; Negative Predictive Value (NPV): 97.7 %. Conclusions: The Random Dot test with Lea symbols does not represent a reliable value to use as single visual screening test for the detection of binocular vision disturbances. The low prevalence (8.31 %) of the condition in this study increases its NPV. This, together with its high specificity, makes the test very efficient when it comes to confirming the likelihood of not having the disorder when a different test has already classified it with normal results.

Keywords

sensitivity, specificity, stereoacuity test, predictive values

Digital Object Identifier (DOI)

https://doi.org/10.19052/sv.1434

Fecha de recepción

17 de agosto de 2012

Fecha de aceptación

17 de septiembre de 2012

Fecha de publicación

January 2012

Licencia Creative Commons

Creative Commons Attribution-Noncommercial-No Derivative Works 4.0 License
This work is licensed under a Creative Commons Attribution-Noncommercial-No Derivative Works 4.0 License.

Tipo de documento

Artículo de investigación

Editorial

Universidad de La Salle. Ediciones Unisalle

Compartir

COinS